Analytical Instruments

TOROIDAL CONDUCTIVITY / TDS TRANSMITTER

CTT 100S

CTT 100S

Product Description

CTT-100S Toroidal Conductivity / TDS Transmitter is a robust, two-wire loop-powered transmitter designed for accurate online conductivity and TDS measurement in demanding industrial applications. It accepts input from a toroidal conductivity electrode, making it ideal for fluids with high conductivity, coating, or fouling tendencies. The measured conductivity signal is transmitted as a stable 4–20 mA output with optional HART communication, ensuring seamless integration with control systems. A backlit LCD display provides clear real-time indication, while auto or manual temperature compensation using RTD PT-100 / PT-1000 improves measurement accuracy. With high accuracy of ±0.25% of full scale, fast response time, and low power consumption, CTT-100S delivers reliable performance. Housed in weatherproof IP65/IP67 or CMRI IIA/IIB flameproof enclosures, it is well suited for chemical processing, wastewater treatment, pulp & paper, and industrial process monitoring applications.

Product Parameters

Configure the transmitter and electronics based on application, environment, and compliance requirements.

Type of Input  From Toroidal Conductivity Electrode
Calibration Range  200 to 2000000 μS/cm
Electronics Transmitter Accuracy  Electronics Transmitter Accuracy +/- 0.25% F. S.
Isolation  1.4 KV between Input, Output & Power Supply
Display  Local display LCD with backlit
Type of Output  4-20 mA DC Output / 4-20 mA with HART (Generic)
Sensor Connection  1) Through BNC connector 2) Through Terminals
Power Supply  24V DC Two Wire Loop Powered
Power Consumption  < 50 mW
Electronic Protection Class  Flame Proof CMRI IIA IIB / Weather Proof IP-65
Response Time  < 500 mSec
Temperature Coefficient  +/- 0.01% per °C
Temperature Compensation  RTD PT-100 / RTD PT-1000
Dimensions  155 mm (H) x 105 mm (W) X 130 mm (D)
Conduit Entry  1) M20X1.5 (F) 2) ½” NPT(F)
Mounting  2” Pipe Mounting
Weight  1kg (Approximate)
Housing  Cast aluminum with epoxy coating NEMA 4X / SS 316
Ambient Conditions  Temperature -20 to 70 °C / Humidity 5 to 95% non condensing at 25°C
Flow Through Assembly (Optional)  SS 316 / PP

Need more technical details?

Download the complete product catalogue for detailed specifications, dimensions, certifications, and application guidance.

Applications

Electronet instruments are widely deployed across multiple industries and utilities for reliable monitoring and control of critical process parameters.

Industry Key Applications
Water & Wastewater Monitoring conductivity and TDS in treatment plants, pipelines, and effluent systems with toroidal sensor technology
Chemical & Petrochemical Measuring solution concentration, process water quality, and chemical dosing accurately using toroidal sensors
Power Generation Conductivity monitoring in boiler feedwater, cooling water, and condensate systems with toroidal measurement
Food & Beverage Monitoring water quality and process liquids for production consistency and safety
Pharmaceuticals & Biotechnology Conductivity/TDS monitoring in process water, bioreactors, and formulation systems with high accuracy
Pulp & Paper Industry Monitoring chemical concentrations and water quality in pulping and bleaching processes
Marine & Shipbuilding Conductivity/TDS measurement in seawater systems, ballast, and cooling circuits
Metals & Mining Monitoring process water and slurry conductivity for process optimization
HVAC & Building Automation Water quality monitoring in cooling towers and chilled water systems
Semiconductor & Electronics Conductivity/TDS monitoring for ultra-pure water and process control

Key Features

Core capabilities and advantages of the product.

  • On-Line Measurement
  • 4-20 mA DC Output / 4-20 mA with HART (Generic)
  • Online Indication through LCD Display
  • Simple Operating Interface
  • 2 Wire System
  • Auto / Manual Temperature Compensation
  • Non-volatile memory retains the program
  • settings during power failures
Send Enquiry